DC Series Low-temperature thermostatic water bath Operation Manual

来源:Ningbo Sjialab Equipment Co., Ltd.时间:2019-07-31关键词:thermostat water oil bath operation manual

 Instrument appearance introduction.

 微信截图_20190731085730.png

 

Features:

1 refrigeration system: With low noise totally enclosed compressor, air cooled with cooling speed, temperature stability is strong, etc.

 

2 control system: Latest proprietary software research and development of homemade temperature control technology, with high standards of Pt100 and all imports of electronic parts, etc.

 

3 circulating system: Consists of inner and outer loop irfpa outer loop slot can be constant temperature liquid, can establish the temperature field, also can be used as cold source to guide groove is cooling liquid into closed experimental container cooling, inner loop can make the liquid in the tank temperature was uniform and stable.

 

4 protection system: Refrigeration is overheated, over-current protection, call control system has overtemperature alarm, can be set overtemperature alarm temperature, the temperature of the upper and lower limits can cut off the load automatically when overtemperature, and other functions.

 

5 High-end LCD control instrument, simple operation, temperature stability is strong, P.I.D can automatically search or artificial adjustment

 

6 Accurate temperature correction function can reach 0.01 ℃, temperature fluctuation degree of up to + / - 0.02 ℃ ~ + / - 0.05 ℃ (depending on the type).

 

 Technique Parameter:

Model

Temperature range

Temperature fluctuation±℃

Working tank volumemm3

Slot opening (mm

Groove depth(mm)

Circulating pump flow(L/min) 

PowerKW

DC-2006

  -20- 100

 0.05

 250x200x150

180x140

150

6

1.0

 

四: Operating  Steps:

1Slot in liquid medium, medium level cannot under working table 20 mm.

2The selection of liquid medium:

AWhen working temperature is below 8 ℃, liquid medium generally choose industrial alcohol.

BWorking temperature 8 ℃, 80 ℃, the pure water is generally selected liquid medium.

CWorking temperature 80 ℃ to 90 ℃, water and oil mixture is generally selected liquid medium.

DWorking temperature at 90 ℃ to 100 ℃, generally choose the oil liquid medium.

3Circulating pump connection:

ACirculating pump in the connection, connect the outlet and inlet hose.

Bexternal circulation pump for external loop connection, connect the outlet hose outside the tank container imports and exports to connect water inlet on the outside of the tank container.

4plugged in, open the total power switch, then open "power" button and "cycle" button. (note: "total power switch on the back of the instrument).

5instrument operation is as follows:

AInstrument button shows:

  shift key    ▲The addend key    ▼The reduction key   SET setting function key

Btemperature setting:

Such as working temperature is higher than the environment temperature, according to the set function keys into the temperature set point. Value bottom flashing, press the shift key before at this time the addend, set up your required working temperature, press set again the function keys and save the setting, the measurement is shown in the groove of the temperature of the liquid medium, then into a state of microcomputer automatic control.

 

CSuch as working temperature below the environment temperature, open "cooling" button (note: the above the liquid temperature 40 ℃) could not be opened, according to the set function keys into the temperature set point. Value bottom, at this time to press the shift key and press reduction, set up your required working temperature, press set again the function keys and save the setting, the measurement is shown in the groove of the temperature of the liquid medium, then into a state of microcomputer automatic control.

 

6temperature and actual temperature measurement, correction:

By setting function button for more than 5 seconds, the window shows the word "PASS", press the add key input password "0001" at this time, again by setting the function keys into the actual revision status, the window shows the word "SC" at this point, press the add key such as high temperature, low press reduction, fix complete set according to the function keys for 5 seconds or so quit again can. Time measured value reaches the working temperature stability experiment or test can be carried out when a period of time.

 

五:Application notice

1Before using slot should be in liquid medium. (note: liquid not contain acid alkali).

2The total power of the power is to be greater than or equal to the instrument, the power must have a good "grounding" device. Using the voltage will be subject to technical parameter table.

3Device should be placed in ventilated dry place, back and sides leave obstacles 300 mm distance

4After use, all switch off and unplug the power plug.

 